Key Technical Specifications:
Input channels & sensor compatibility
-
3 wired input channels: 1 for Pt100 sensors + 2 for thermocouples (TC) Type K, T, J, S.
-
Support for up to 3 additional wireless temperature sensors (radio probes) for a total of up to 6 channels (3 wired + 3 wireless).
-
Compatible probe types: Pt100 immersion/penetration, TC K (NiCr-Ni), TC T (Cu-CuNi), TC J (Fe-CuNi), TC S (Pt10Rh-Pt).
Measuring ranges, accuracy & resolution (typical)
-
Pt100 sensor input: Measuring range: –200 °C to +800 °C. Accuracy: ±0.2 °C (for –100 to +199.9 °C) + ±0.2 % of measurement value for remaining range. Resolution: 0.05 °C.
-
Thermocouple Type K: Measuring range: –200 °C to +1 370 °C. Accuracy: ±0.3 °C (for –60 to +60 °C) + ±(0.2 °C + 0.3 % of mv) for remaining range. Resolution: 0.1 °C.
-
Thermocouple Type T: Measuring range: –200 °C to +400 °C. Accuracy: ±0.3 °C (for –60 to +60 °C) + ±(0.2 °C + 0.3 % of mv). Resolution: 0.1 °C.
-
Thermocouple Type J: Measuring range: –200 °C to +1 000 °C. Accuracy: ±0.3 °C (for –60 to +60 °C) + ±(0.2 °C + 0.3 % of mv). Resolution: 0.1 °C.
-
Thermocouple Type S: Measuring range: 0 °C to +1 760 °C. Accuracy: ±1 °C (for full range). Resolution: 1 °C.
-
When using the optional high-precision Pt100 immersion/penetration probe (e.g., probe 0614 0235) the system accuracy can reach ≈ 0.05 °C with resolution of 0.001 °C.
Memory & data logging
-
Internal memory: Up to 10 000 measurement values can be stored.
-
Supports up to 99 measurement sites and multiple user profiles.
-
Logging (memory cycle) selectable from 0.5 seconds up to 24 hours.
- PC-software included with USB data cable for download, analysis (tabular/graphical) and documentation of measured data.
General / housing / environmental
-
Housing dimensions: 220 × 74 × 46 mm. Weight: approx. 428 g.
-
Operating temperature: –20 °C to +50 °C. Storage temperature: –30 °C to +70 °C.
-
Protection class: IP65 (dust tight and protected against water jets).
-
Housing material: ABS / TPE / Metal.
-
Power supply: AA (mignon) alkaline batteries. Typical battery life (with TC probe K/T): ~200 h.
-
Standard compliance: EN 13485 and HACCP-International certified (for food/pharma industries).